V. Using the Autosalinometer to Check Data Accuracy
In order to cross-check the salinity data coming from the SEASOFT program, take samples from casts and find their salinity using the Guildline Autosalinometers located in the closet in the Bio Lab. This is often referred to as "running salts." The salinometer is a very sensitive instrument, so it is imperative to follow the guidelines below. Also, the instruction book for the instrument is located in the salinometer room and provides some more information. Please see Appendix E for notes on sampling techniques.
Occasionally, it may be necessary to use the Portasal, a portable salinometer that is easier to operate than the Autosal, but provides lower accuracy. A complete procedure is located in Appendix F. Highlighted differences between the Autosal and the Portasal are located toward the end of this section.
About the Autosalinometer
An image of the Autosal and Portasal can be found in Appendix I.
The Guildline Autosalinometer (Autosal) is very temperature sensitive. The temperature reading on the machine must be within four degrees of the room temperature (on the warmer side-the best setting is typically 24 C). The samples should be at room temperature; put them in the closet at least 12 hours before running them. The unit has an easier time warming samples than it does cooling them. Keep the door to the salinometer room closed to keep the temperature stable. If there's a problem keeping the temperature stable, close the door between the hall and the Bio Lab. Temperature control is VERY important.
It is very easy to contaminate a sample. To prevent this, wear rubber or laytex gloves, and don't touch anything that may go into the sample or the Autosal. Use a new kimwipe to wipe down the sample tube after each sample is run.
Make sure there are no bubbles in the conductivity cell. This affects readings severely.
Before running a sample, shake it very well to mix any stratified layers. Also, wipe off the mouth of the sample bottle very carefully, removing any salt crystals.
- gonewiththeflow > Autosalinometers
-
Autosalinometers
下载该文档 文档格式:DOC 更新时间:2010-06-01 下载次数:0 点击次数:1
文档基本属性 文档语言: English 文档格式: doc 文档作者: Marc Pomeroy 关键词: 主题: 备注: 点击这里显示更多文档属性 经理: 单位: 分类: 创建时间: 上次保存者: 修订次数: 编辑时间: 文档创建者: 修订: 加密标识: 幻灯片: 段落数: 字节数: 备注: 演示格式: 上次保存时间:
- 下载地址 (推荐使用迅雷下载地址,速度快,支持断点续传)
-
DOC格式下载
- 更多文档...
-
上一篇:up-to-the-minute
下一篇:ppeckol@emailsmithedu
点击查看更多关于gonewiththeflow的相关文档
- 您可能感兴趣的
- gonewiththewind gonewiththesin gowiththeflow flowwiththetide gonewiththegun gonewiththesi gonewithwind gonewithsin gonewithleaves gonewithyesterday
- 大家在找
-
- · 2011年高考作文范文
- · 三菱fx系列解密软件
- · 洗瓶机设计模板
- · 2012影像技师考试
- · 面盆防臭下水器
- · 五十铃皮卡柴油发动机
- · 死神vs火影1.4小游戏
- · 攻宠受bl小说打包下载
- · 艺术欣赏对职校学生的作用
- · 优秀论文获奖证书模板
- · 2010年nba全明星mvp
- · 抗生素药物分类
- · 成都北大青鸟it培训
- · 清华大学化工系
- · 不死冥王电子书下载
- · 城堡战争1.13指令
- · 本科护理论文范文
- · 西门子冰箱质量怎么样
- · cs新boss
- · 陈式83式太极拳谱
- · 联想a60自带软件下载
- · 高中美术说课稿范文
- · 深圳龙华清湖招临时工
- · 发那科系统
- · 半导体猜一字
- · 电焊技能视频下载
- · 免费下载exel2003
- · 高三金鼎一试卷联考卷一
- · 江都宾馆招聘客房服务
- · 花搂恋歌舞蹈视频
- 赞助商链接